Using N8480 Series Power Sensors

N8480 Series Sensors Connection Requirements

Table 8-15 N8480 Series connection requirements

200
Sensor
N8481A
These power sensors connect directly to the POWER REF.
N8481H
N8482A
N8482H
N8481B
These power sensors are configured with an attenuator. Prior to
N8482B
calibration this attenuator must be removed. Replace the attenuator
before making measurements.
This power sensor requires an APC 3.5 (f) to 50 Ω (m) N-Type adapter
N8485A
(08485-60005) to connect to the POWER REF.Remove this adapter before
making measurements.
N8486AR
These waveguide power sensors have two connectors. Use the
N8486AQ
N-Type connector to calibrate the power meter.
This sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
N8487A
N8488A
(08487-60001) to connect to the POWER REF. Remove this adapter
before making measurements.
Connection Requirements
N1911A/1912A P-Series Power Meters User's Guide