CCDF
Expanded CCDF Table
Press

Figure 9-148CCDF table in expanded display

In expanded format, you are allowed to query the power (dB) and the
probability (%) based on your input. You can also vary the sample count of
the CCDF table.
To set the sample count, press
can also set the mode to single or continuous. In single mode, the graph
and table readings will stop refreshing after the desired elapsed count is
acquired. In continuous mode, the graph and table readings will reset
after each cycle of elapsed count.
The mode can be set by pressing
The minimum elapsed count is 100 M, while the maximum is 10 G.
N O T E
220
once for an expanded view of the CCDF table.
Statistical Settings
N1911A/1912A P-Series Power Meters User's Guide
then
Mode Single Cont
.
Counts 100.0M
. You