Maximum dynamic range is related to the maximum sensor bandwidth. Please refer to the
N O T E
documentation supplied with your E-Series E9320 power sensor for specification
information.
Figure 4- 101
E9327A power sensors

Figure 4-101 Bandwidth Filter Shapes

N1911A/1912A P-Series Power Meters User's Guide
shows the filter shapes associated with the E9323A and
Amplitude
0
–3.5 dB
Approx
300 kHz 1.5 MHz
Meter B/W
Low
Setting
Med
High
Off
Using E9320 E-Series Power Sensors
E9323A and E9327A sensors
Frequency
5 MHz