Figure 2-35 Bandwidth filter shapes

When video bandwidth is set to Off, it removes all digital signal
conditioning. This provides approximately 3 dB roll off at the maximum
sensor's bandwidth and is best suited for accurate trace capture,
minimizing overshoot, and removing any ringing effects caused by the
sharp cut- off filters used in the Low, Med and High settings.
shows the filter shapes associated with the P- Series wideband power
sensors.
Procedure
Set the video bandwidth as follows:
1 Press
2 Use the
3 Press
N1911A/1912A P-Series Power Meters User's Guide
Amplitude
0
–3.5 dB
Approx
5 MHz
Meter B/W
Low
Setting
Med
High
Off
. On dual channel meters select the required channel.
and
to display the Video B/W pop- up.
P-Series Sensors
15 MHz
30 MHz
keys to select the Video B/W setting field.
General Power Meter Functions
Frequency
2
Figure 2- 35