Zeroing And Calibrating The P-Series Sensor; Figure 2-87 Zeroing Pop-Up - Agilent Technologies N1911A User Manual

P-series power meters
Hide thumbs Also See for N1911A:
Table of Contents

Advertisement

Zeroing and Calibrating the P-Series Sensor

This section describes how to zero and calibrate the power meter when
using a P- Series wideband power sensor.
Zeroing
Zeroing adjusts the power meter for a zero power reading on each power
meter channel and P- Series sensor combination. This is achieved without
removing it from a power source. The power meter can be set to
automatically zero on- the- fly while you are using it. Alternatively, you can
perform a zero at your preference. During zeroing a wait symbol is
displayed.

Figure 2-87 Zeroing pop-up

When the P- Series power sensor is initially connected to the power meter,
it automatically performs a Zero and Calibration routine.
When to Zero?
Zeroing of the power meter is recommended:
• When a 5
• Every 24 hours.
• Prior to measuring low level signals. For example, 10 dB above the
Automatic Zero
The automatic zero can be enabled as follows:
N1911A/1912A P-Series Power Meters User's Guide
o
C change in temperature occurs.
lowest specified power for your power sensor.
General Power Meter Functions
2
107

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

1912aN1912a

Table of Contents