General Power Meter Functions
2
Offset Indicator

Figure 2-26 Channel offset indicator

The Ofs symbol is not displayed when the associated measurement is displayed in Dual
N O T E
Numeric or Analog format.
Setting Display Offsets
This gain or loss is applied to the measured power after any channel
offsets or mathematical functions have been included.
Offsets are entered in dB and the range of values is –100 dB to +100 dB.
A positive value compensates for a loss, and a negative value compensates
for a gain.
Procedure
Enter a display offset on the currently selected window:
1 Press
to display the Measurement Setup screen.
2 Select the window you wish to set the offset value on by pressing the
Meas Select
key. The currently selected window/measurement is
displayed.
3 Use the
and
keys to highlight the Offset setting field.
4 Press
to check the Offset setting field.
N1911A/1912A P-Series Power Meters User's Guide
41