Setting Measurement Channel Gates

This feature is only available when a P-Series or an E-Series E9320 power sensor is
N O T E
connected.
A system of gates, controlled by and referenced to a trigger point, is used
to obtain measurement data from a captured trace. The trace data within
each gate period is subsequently used for the individual measurement
calculations. Up to 4 gates can be set up for each channel.
shows an example of 4 gates setup to perform the following measurements
simultaneously:
• Average power level of the pulse:
• Average "off" power level ahead of the pulse:
• Peak to average ratio:
• Pulse drop:

Figure 2-37 Example of measurement gates

N1911A/1912A P-Series Power Meters User's Guide
Gate 1, average measurement
Gate 2, average measurement
Gate 1, peak- to- average measurement
Gate 3, average measurement, minus Gate 4, average measurement
Gate 3
Trigger
Gate 2
General Power Meter Functions
Gate 4
Gate 1
2
Figure 2- 37
57