c. Press the OPTIONS or PRIME keys if either of these functions is desired at this time.
2. REPEAT Rx: To repeat the present therapy and add another IV container with the same
prescription, press the YES/ENTER key at this eld. If more than one therapy is programmed,
a Select
erapy Screen will list all preprogrammed therapies. Select which therapy to repeat by
moving the cursor to the therapy of choice and pressing the YES/ENTER key.
e pump will then display the program screen with the last programmed values of the therapy
and the cursor will automatically advance through the prescription menus. You can stop the auto-
advance function by pressing any key (except ON/OFF) and then complete the con rmation of
the prescription manually by pressing the YES key at the remaining elds.
Note:
e previously programmed AIL setting is also retained
When the program is accepted, the display returns to the above Run Options Screen. If the
current therapy is still in progress when the "REPEAT Rx" function is selected, the following
screen will appear:
a.
is screen is a reminder that the infusion is not complete. If the therapy is to be terminated
and a new IV container added, press YES/ENTER to go to the programming screen, con rm
the program and proceed with the repeat infusion.
b.
is screen is a reminder that the infusion is not complete. If the incomplete therapy is to be
continued, press NO/CHANGE and the Pause Menu Screen will be displayed, then press YES/
ENTER and the Run Options Screen will be displayed. To resume the un nished therapy, press
RUN.
c. If the therapy is to be terminated and a new IV container added, press YES/ENTER to repeat
the infusion.
e following screen will then appear, asking whether or not you wish to maintain
the current Bolus or MedLMTS™ schedule.
H a nging a N e w Cont a ine r
8 7