b. If the incomplete therapy is to be continued, press NO and the Pause Menu Screen will
be displayed, then press YES and the Run Options Screen will be displayed. To resume the
un nished therapy, press RUN.
3. NEW PROGRAM: To program an entirely new prescription or di erent therapy, press the
YES/ENTER key at this eld (the ability to change these parameters may be controlled by the
lock setting).
e display will return to the Select
is enabled, or to the Pre-Prescription Screen of the single therapy mode. If the current container
has not completely infused, the following screen will appear:
a.
is screen is a reminder that the infusion is not complete. If you want to return to the
current container and nish infusing it, press the NO/CHANGE key to return to the Pause
Menu Screen.
b. To abort the current container and enter a new program, press the YES/ENTER key. If a
prescription has previously been entered for the therapy mode you wish to program, the
following screen will appear:
c. If you want to save the current prescription, press the NO/CHANGE key to return to the
Pause Menu screen. If you wish to erase the current prescription and enter a new program,
press the YES/ENTER key to proceed. All previous prescription values will be erased.
Note: Selecting "NEW PROGRAM" will also reset the OPTIONS selected to their default
settings except for the LOCK and AUDIO settings which must be changed manually.
When the infusion is complete, the audio alarm will beep, and the following screen will display:
Note: To temporarily silence the alarm, press the SILENCE key.