Moog Curlin 6000 CMS IOD User Manual page 180

Ambulatory infusion system

Advertisement

Power Requirements
ITEM
Operating Time with Full Charge
Batteries
Internal Memory Battery
Display
Flow Rates
KVO Rates
Volume Accuracy
Administration Sets
Fill Volume for Administration Sets
Up Occlusion Detection Time
Microbore tubing with occlusion below
spike
2 C-cell Single-Use Batteries
AC Power
SPECIFICATION
Approximately 10 hours at a rate of 400 mL/hr
Approximately 30 hours at a rate of 125 mL/hr
Approximately 85 hours at a rate of 2 mL/hr
5 years life
Graphic LCD, 4 text elds, 16 characters per
eld
0.1 to 400 mL/hr
0 to 10 mL/hr. Can be set at "0" in
Continuous, TPN, Variable and Intermittent
erapies
± 5% for volume delivery >1 mL under the
following conditions:
• Temperature 55° to 90° F
• Back pressure to 300 mmHg (5.8 PSI)
±10% for volume delivery >1 mL under
the following conditions:
• Temperature 45° to 55°F
• Temperature 90° to 104°F
• Back pressure to 750 mmHg (14.5 PSI)
e maximum volume that may be infused
under Single Fault Conditions is 0.8 mL.
Recommended administration set usage not to
exceed 72* hours.
*follow standards of practice for set changes
Proprietary Moog Curlin Infusion non-DE-
HP, PVC administration sets available in
di erent con gurations.
Non ltered microbore sets ± 4 mL.
See individual administration set label for
speci c ll volume
Rate
0.1 mL/hr
1 mL/hr
10 mL/hr
50 mL/hr
1 6 5
Time to Alarm
<25 minutes
< 4 minutes
<20 seconds
<10 seconds