Resolution:
Install the appropriate Moog Curlin Infusion administration set for the desired
therapy. Use only Moog Curlin Infusion administration sets. See Chapter 1,
gures 1.9 and 1.10 of the User's Manual for proper installation of the
administration set.
LCD Display:
"***ALERT *** SYSTEM CLOCK NOT SET"
Alert:
System Clock Not Set.
Cause:
e clock for the pump is not set
Resolution:
Use the BIOMED SETUP menu "Date and Time" option to set the date and
time.
LCD Display:
"***ALERT *** THERAPY TIME INVALID"
Alert:
Invalid therapy time calculated.
Cause:
While programming or modifying an Intermittent therapy, the combination
of dose duration and dose frequency result in a prescription with a duration
greater than 288:00 HH:MM.
Resolution:
Recheck data entry.
limit.
LCD Display:
"***ALERT *** TIME LESS THAN RAMP TIME"
Alert:
Unable to enter a TPN therapy time that is less than the ramp times.
Cause:
Time value entered for TPN infusion is less than ramp times.
Resolution:
Recheck data entry. TPN infusion time cannot be less than speci ed ramp
times.
LCD Display:
"***ALERT *** VALUE PREVENTS A SUBSEQUENT DOSE"
Alert:
Value entered will prevent the ability to program any more doses.
Cause:
e value entered for the "Variable Dose Amt" or "Time eld" will cause no
remaining amount or time for a subsequent.
Resolution:
Recheck data entry. If this is a new program, reprogramming of the therapy
will be required if the error is in a previously programmed dose.
If the therapy is in progress or this is a repeat program that has not started and
the error is in a previously programmed dose, turn the pump o and then on.
Choosing to resume or repeat the therapy will display the last prescription that
was completely programmed.
LCD Display:
"***ALERT *** VALUE
Alert:
Value entered is above the Library Limit.
Cause:
e value entered is out of range of the soft limit setting(s) for a downloaded
therapy.
Resolution:
e screen will ask the user to override the soft limit value. Pressing YES will
accept the new value.
e changes to the eld must comply with the total time
LIMIT"