1. RESUME: To resume the therapy in progress from where it left o , press the YES/ENTER
key at this eld.
therapy.
Note: If you wish to hang a new container, move the cursor to "REPEAT Rx" (see following
section).
a. Press the RUN key to resume the therapy.
b. Press the NO key to go back to the programming screen to make changes in the therapy
program (this feature may be blocked by a security lock, thus not allowing changes to the
program unless an access code is given). Also, remember that no changes are allowed in the
rst prescription menu if an infusion is in progress.
c. Press the OPTIONS or PRIME keys if either of these functions is desired at this time.
2. REPEAT Rx: To repeat the present therapy and add another IV container with the same
prescription, press the YES/ENTER key at this eld. If more than one therapy is programmed,
a Select
erapy Screen will list all preprogrammed therapies. Select which therapy to repeat by
moving the cursor to the therapy of choice and pressing the YES/ENTER key.
e pump will then display the program screen with the last programmed values of the therapy
and the cursor will automatically advance through the prescription menus. You can stop the auto-
advance function by pressing any key (except ON/OFF) and then complete the con rmation of
the prescription manually by pressing the YES key at the remaining elds.
Note:
e previously programmed AIL setting is also retained
When the program is accepted, the display returns to the above Run Options Screen. If the
current therapy is still in progress when the "REPEAT Rx" function is selected, the following
screen will appear:
a.
is screen is a reminder that the infusion is not complete. If the therapy is to be
terminated and a new IV container added, press YES/ENTER to go to the programming
screen, con rm the program and proceed with the repeat infusion.
b. If the incomplete therapy is to be continued, press NO and the Pause Menu screen will
be displayed, then press YES and the Run Options Screen will be displayed. To resume the
un nished therapy, press RUN.
Re sum ing t he Curre nt Cont a ine r
e pump will return to the Run Options Screen, and you can resume the
H a nging a N e w Cont a ine r
7 0