is message alerts you to the Air-In-Line Sensor being set to "OFF."
answering "yes" or "no" to whether the administration set being used has an air-eliminating lter.
If you press the YES key to indicate that there is an air-eliminating lter in use, the Air-In-Line
Sensor will be completely disabled and the pump will start the infusion.
If you press the NO key to indicate a lter is not in use, the Air-In-Line Sensor will be set to
detect the presence of any contiguous 2-milliliter air in the line.
then brie y appear on the screen. If the alert occurred at the Run Options Screen, the pump
will start the infusion.
A i r - I n - L i n e
Also, when the Air-In-Line Sensor is completely disabled and the infusion goes to an "Infusion
Complete" or "Empty Bag" status, the Air Sensor setting will automatically change to 2 mL.
6. NXT DOSE (Next Dose): Allows the time of the next dose to be changed.
remaining-until-the-next-dose is displayed. Use the numeric keys to change the time for the start
of the next dose. If correct, press the YES/ENTER key to accept, and move to the next eld.
range for this eld is 0:00 to 24:00 HH:MM.
Note: If there is a dose currently in progress and the NXT DOSE feature is used, the current
dose will be terminated and the cycle will continue at the KVO rate until the time for the
new next dose.
7. AUDIO (Audio Volume):
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting).
When you select a number, the pump will beep at that audio level. Press the NO/CHANGE key
and enter a new setting (1 - 9). Repeat until the desired setting is reached, and then press YES/
ENTER to select it and move the cursor to the next eld.
8. PWR CK? (Power Check):
display gives the active power source of the pump, whether internal batteries or external power.
If the active source is the batteries, a bar graph will indicate approximately how much charge
remains in the batteries. To access this information, use the NO/CHANGE key to change from
the default setting of "OFF" to "ON," and press YES/ENTER.
display for ve seconds and return to the Options Menu.
9. ACCEPT?: If the changes made in the Options Menu are acceptable, press the YES/ENTER
key. If the changes are not acceptable, use the arrow keys to cursor up to any eld for
corrections, or abort all changes made in this menu by pressing the NO key and return to the
previous screen.
S e n s o r
S e t
a t
2 m L
is feature allows you to adjust the audio volume of the pump's
is feature allows the user to recheck the battery charge.
1 1 2
e message will require
e alert message below will
A l e r t
S c r e e n
e current time-
e Power Source Screen will
e
e