Beaker with clean, fresh water or IV container.
Note: If the administration set contains a lter, sterile saline or sterile water should be used.
is procedure does not replace the recommended preventive maintenance.
receive preventive maintenance by a quali ed technician annually.
1. CLEAN AND INSPECT PUMP FOR DAMAGE:
a. Clean the pump with a soft cloth dampened with a designated antiseptic, disinfectant
solution or a solution of household bleach diluted 9:1 with water. Be sure to clean all
surfaces, including the pumping chamber.
b. Inspect pump for damage. Check all case surfaces for cracks. Open the pumping chamber
door. Examine the nger seal to con rm that no cuts or tears exist. Con rm that the door
is well aligned by closing it with no unreasonable interference. Verify that the power and
Bolus/Data connectors are securely mounted on the case (no loose nuts). Verify that all
labels are legible and in good condition. Check that Keypad is securely attached to case and
that no corners are pealing up. Verify that the battery oor is in place and that the battery
contacts in the battery compartment are not deformed.
2. VOLUMETRIC DELIVERY TEST:
Refer to the picture below when executing the following instructions.
3.
Volumetric Delivery Test:
Maintenance Between Patients
1 6 1
e pump should