Resolution:
Review the User's Manual for program restrictions or if lock level restricted,
change lock setting to access area you wish to modify.
LCD Display:
"***ALERT *** NEW DOSE IS LESS THAN CURRENT DOSE"
Alert:
Unable to program a lower number of doses than number of doses already
infused
Cause:
While the Variable or Intermittent therapy is in progress, a lower number of
doses to be delivered is entered than the number already infused.
Resolution:
Wait until current therapy in progress is complete, or PAUSE pump and,
using the "REPEAT Rx" function, modify the number of doses.
Note: Selecting to repeat the therapy in progress will terminate the current
therapy.
LCD Display:
"***ALERT *** NO BOLUS DOSE DEFINED"
Alert:
Unable to enter bolus interval or # of bolus/hr because no bolus value has been
programmed.
Cause:
Entry of a value other than zero for the Bolus Interval or #Bolus per hour eld
when no bolus value has been programmed.
Resolution:
Enter bolus value before entering bolus interval or # of boluses/hr.
LCD Display:
"***ALERT *** NO KEYPAD ACTIVITY"
Alert:
No Keypad Activity.
Cause:
e pump has been displaying a non-running screen for two minutes without
any keys being pressed.
no key presses.
Resolution:
Complete the activity for the displayed menu.
o , however, all programming that was not complete will be discarded.
If the pump was running and the Options menu displayed when the alert
occurred, the run screen for the therapy will be displayed when the alert is
cleared. Any changes to the Options menu are discarded and will have to be re-
entered.
LCD Display:
"***ALERT *** NO THERAPY ENABLED"
Alert:
No
erapy Enabled.
Cause:
Attempting to accept the Enable
At least one therapy must be enabled or "ON" for menu to be accepted.
Resolution:
Enable the therapy or therapies that are to be active for the pump.
LCD Display:
"***ALERT *** NOT ALLOWED, DOSE IN PROGRESS"
Alert:
Change is not allowed while a dose is in progress.
e alert will recur every 2 minutes as long as there are
e pump may also be turned
erapy menu with all therapies set to "OFF."
1 4 1
Need help?
Do you have a question about the Curlin 6000 CMS IOD and is the answer not in the manual?
Questions and answers