If the therapy will not be repeated, press PAUSE key and turn the pump o
by pressing the OFF key.
LCD Display:
"STUCK BOLUS KEY"
Alarm:
Bolus Key is stuck.
Cause:
e Remote Bolus Cord is not functioning correctly.
Resolution:
Replace the Remote Bolus Cord.
LCD Display:
"***ALARM *** LOW MOTOR SUPPLY"
Alarm:
Low Motor Supply Voltage.
Cause:
e voltage to the Motor Supply is too low.
Resolution:
Connect the AC adaptor or replace the batteries.
LCD Display:
"***ALARM *** MOTOR STALLED"
Alarm:
Motor Stalled
Cause:
e motor is not turning at the expected rate.
Resolution:
Connect the AC adaptor or replace the batteries and resume the infusion.
LCD Display:
"***ALARM *** INVALID UNITS FOR PUMP CONFIGURATION"
Alarm:
Invalid Units for Pump Con guration.
Cause:
e Library Protocol units are not valid for the pump con guration
Resolution:
Select a di erent Library Protocol or revise the Library Protocol.
LCD Display:
"***ALARM *** INVALID ROUTE FOR PUMP CONFIGURATION"
Alarm:
Invalid Route for Pump Con guration.
Cause:
e Library Protocol Route is not valid for the pump con guration.
Resolution:
Select a di erent Library Protocol or revise the Library Protocol.