move on to the next.
2. Amt TBI (Amount To Be Infused): Use the numeric keys to enter the actual amount of
infusion to be delivered excluding any KVO amounts.
prescription menu will automatically appear in this eld, whether mL, mg, or mcg. When the
amount is correctly entered, use the YES/ENTER key to accept this eld and to move to
the next. (If the Amt TBI entered is greater than Bag Volume, the pump will beep once. A
message of "AMT TBI GREATER THAN BAG VOL" will appear to remind you to
recheck and change the number before the YES/ENTER key will accept the value and move to
the next eld.)
e range for this eld is 1 to 9999.
3. RATE:
e Moog Curlin Infusion pump can deliver from 0.1 mL to 400 mL per hour. Use
the numeric keys to enter the prescribed rate. If the time of the infusion is entered rst, the
pump will automatically calculate the rate. When this eld is correct, use the YES/ENTER key
to accept the rate, and move to the next eld.
4. TIME: Displays the time required to complete the infusion. When the pump is given the
Amount To Be Infused and the RATE at which to infuse, it will automatically calculate the Time
and display it here. If the time is not acceptable, press the NO key to edit this eld, and enter
the new TIME with the numeric keys. (Remember, by changing the time of the infusion, a new
RATE will be calculated and will have to be accepted before exiting this menu.
Be Infused" will remain constant.) When the TIME eld is satisfactory, press the YES/ENTER
key and move to the nal eld.
Note:
e time calculated by the pump may be less than a minute; however, one minute is the
minimum entry and display value.
5. KVO Rate (Keep Vein Open Rate):
the rate at which uid will be delivered in a "Keep Open" state.
0.1 mL/hour, can be set from 0 to 10 mL/hour, but cannot be set at a rate to exceed the therapy-
operating rate. When the desired KVO rate has been entered, press the YES/ENTER key to
accept and move the cursor to the next eld.
6. DONE?: If any changes are to be made, use the up or down arrow keys to locate the eld to
be changed, use the numeric keys to enter the corrected information, and press YES/ENTER to
accept the new data. If all the elds are acceptable, move the cursor to the "DONE?" eld and
press the YES/ENTER key to move on to the next screen.
Before the infusion is started, it is important to consider a few remaining issues, such as
I M PORTAN T
I N FORM AT I ON
e range for this eld is 1 to 9999.
e range for this eld is 0:01 to 288:00 HH:MM.
is eld requires data entry using the numeric keys to set
checking the administration set to be sure it was primed or purged of
all air, or perhaps the pump may need to be put into a security lock
setting for additional safety.
e next screen that appears when the programming is complete o ers
6 2
e "Units" selected in the
e pump defaults to a rate of
e "Amount To

Advertisement

loading
Need help?

Need help?

Do you have a question about the Curlin 6000 CMS IOD and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

This manual is also suitable for:

Curlin 6000 cms6000 cms6000 cms iod