5. NEXT?: If there are any changes to be made, use the arrow keys to locate the eld to be
changed, use the numeric keys to enter the corrected information, and press YES/ENTER to
accept the new data. If all elds are acceptable, move the cursor to the "NEXT?" eld and press
the YES/ENTER key to accept and move on to the next dose programming screen. When the
total number of doses is programmed, the pump will display the Variable Summary Screen.
e summary screen indicates the total number of doses programmed and the total time and
volume required.
e elds on this screen cannot be modi ed.
1. TOT DOSES:
e eld displays the total number of doses programmed. If correct, press the
YES/ENTER key to accept and move to the next eld.
2. TOT Time:
e pump calculates the total time required to deliver the total number of doses.
Press the YES/ENTER key to accept and move to the next eld.
3. REQ VOL (Required Volume):
doses. Press the YES/ENTER key to accept and move to the next eld.
4. DONE?: If any of the information is NOT acceptable, press the YES key and then
the NO key at the Run Options Screen.
Prescription screens to allow changes to be made. When all elds are acceptable, move the
cursor to the "DONE?" eld and press the YES/ENTER key to move on to the next screen.
Before the infusion is started, it is important to consider a few
I M PORTAN T
remaining issues, such as checking the administration set to be sure
I N FORM AT I ON
it was primed or purged of all air, or perhaps the pump may need to
be put into a security lock setting for additional safety.
e following screen o ers four choices:
Note that you do not need to answer yes or no questions or enter data with the numeric keys as
previously requested.
is screen requires only that you press the corresponding key on the keypad
to begin the requested function.
V a r i a b l e
S u m m a r y
e pump calculates the total volume required to deliver the
e display will return to the Pre-Prescription and
R u n
O p t i o n s
e following explains each selection beginning with Prime:
1 2 3
S c r e e n
S c r e e n
Need help?
Do you have a question about the Curlin 6000 CMS IOD and is the answer not in the manual?
Questions and answers