Download Print this page

Moog Curlin 6000 CMS IOD User Manual page 154

Ambulatory infusion system

Advertisement

Cause:
Modi cation to the current eld is not allowed while the dose is in progress.
Resolution:
Wait until the dose in progress completes before modifying eld.
LCD Display:
"***ALERT *** NOT in PCA. NO BOLUS ALLOWED"
Alert:
Bolus requested when therapy is not PCA.
Cause:
Remote Bolus button or BOLUS key is pressed when not in PCA therapy.
Resolution:
Remove the remote bolus cord from the pump if connected. Refrain from
pressing the BOLUS key.
LCD Display:
"***ALERT *** NUMBER OF DOSES INVALID"
Alert:
Number of doses entered exceeds maximum for therapy.
Cause:
e number of doses per bag in Intermittent therapy exceeds the allowable
maximum.
Resolution:
Recheck data entry.
to the maximum for the therapy.
LCD Display:
"***ALERT *** PERFORM ROUTINE MAINTENANCE"
Alert:
Maintenance Date (displayed only if date has arrived when pump is turned on)
Cause:
Current Date is beyond the Maintenance Date.
Resolution:
Return pump for routine maintenance.
LCD Display:
"***ALERT *** RAMP INVALID FOR AMT/RATE"
Alert:
Ramp time entered is invalid for amt/rate already programmed.
Cause:
Ramp value entered for TPN is unacceptable given the desired rate and amount
programmed.
Resolution:
Recheck all data entered for prescription. If programming a new TPN therapy
("NEW PROGRAM" selected) and the ramp time is a requirement, enter the
ramp time(s) and total time rst followed by the next governing parameter (rate
or volume to be infused).
LCD Display:
"***ALERT *** RATE EXCEEDS DELTA RATE/DELTA TIME"
Alert:
Rate entered exceeds the delta time limits.
Cause:
e rate entered falls outside of the allowable range for the delta rate / delta time
limits programmed.
Resolution:
Recheck data entry or wait more time elapses then attempt to change the rate.
Rate selected cannot exceed the range placed on MedLMTS
LCD Display:
"***ALERT *** RUNNING, PAUSE PUMP PRIOR TO PRESSING OFF"
e number of doses requested must be less than or equal
1 4 2
.