Download Print this page

Moog Curlin 6000 CMS IOD User Manual page 59

Ambulatory infusion system

Advertisement

When you select "YES," all patient data clears and the display returns to the BIOMED SETUP
Menu. Selecting "NO" aborts the feature and also returns to the menu.
13. CHANGE CONTRAST:
is no longer changeable with this option.
14. CHANGE DATE AND TIME: When this eld is selected, the screen on the left appears:
C h a n g e
is screen is used to set the system Real Time Clock in the pump and will change the date and
time displayed on the Welcome Screen.
Note:
e time must be reset to accommodate daylight savings time changes and also at any
time the pump is moved into a di erent time zone.
a. Time Format: You have the option of selecting twelve-hour or twenty-four-hour time
format. Press the NO key until the correct selection appears on the eld and then press
the YES/ENTER key to accept the selection and move the cursor to the next eld.
b. AM/PM:
e second eld allows you to select AM or PM when the twelve-hour time
format is selected in the previous eld. If the twenty-four-hour format was selected, this
eld will not appear. Press the NO key until the correct selection appears on the eld,
and then press the YES/ENTER key to choose the selection and move the cursor to the
next eld.
c. Time: Use the numeric keys to enter the correct time and then press the YES/
ENTER key to accept and move the cursor to the next eld.
d. Using the numeric keys, enter the correct mm/dd/yy, using two digits for each (e.g., if
the month is January, enter 01; if it is a single digit date, use the number preceded by a 0
and then enter the last two digits of the year code, 00, 01, etc.). When the date is correct,
press the YES/ENTER key to accept and move to the next eld.
e. A nal prompt will be given to "ACCEPT? YES/NO." If any data needs to be
corrected, use the arrow keys to locate the correct eld, press NO to change the data,
enter the correction(s), and then press YES/ENTER again to accept the correction.
Move the cursor back to the "ACCEPT? YES" eld and press YES/ENTER.
enters the new time and date and moves to the next eld in the BIOMED SETUP
Menu. Pressing the NO key on this eld aborts the change date and time selection and
returns to the BIOMED SETUP Menu.
e contrast on the display screen is automatically optimized and
Cha nging Dat e a nd T im e
D a t e
a n d
T i m e
5 0
S c r e e n
is action

Advertisement

loading
Need help?

Need help?

Do you have a question about the Curlin 6000 CMS IOD and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Curlin 6000 cms6000 cms6000 cms iod