will change to "ON." Now, when you press the YES/ENTER to accept, the infusion start time
screen will appear:
e time format (12 hr AM/PM or 24 hr Military) selected in the BIOMED SETUP Menu
will apply in this table; and, if Military was selected, the AM/PM eld will not appear here. If
the entire delay start screen is to be aborted, press the NO key at the "ACCEPT" eld. When
the start time and date are entered and the data is accepted after the YES/ENTER key is pressed,
the infusion is set to start at the programmed time.
the Pre-Prescription Menu.
After the entire prescription program is accepted, if the pump is turned o and back on, press
YES at "PROGRAM" and YES to "REPEAT Rx."
the program. No changes will be allowed in the Pre-Prescription Menu during this time, thus
maintaining the originally programmed Delay Start time. To change the Delay Start time, press
the NO/CHANGE key when the Run Options Screen is displayed:
is will allow the clinician to go back to review the program and at this point a change can be
Note: If the pump is not connected to the patient's access site at this time, the patient or caregiver
should be taught proper techniques to access the infusion site before the scheduled start time.
When the RUN key is pressed, the green and yellow LED lights will ash to indicate that the
pump will automatically start when the delay start countdown period is up. During the delay
start time, the following screen appears:
If the KVO rate is set to "0," the pump will beep ten minutes prior to the beginning of the
therapy start time. To stop the audible beep, press the SILENCE key.
I M PORTAN T
I N FORM AT I ON
4. NEXT?: If there are any changes to be made, use the up or down arrow keys to locate the eld
to be changed, use the numeric keys to enter the corrected information and press
R u n
O p t i o n s
made to the delay start time.
During the delay start period, if the KVO rate is set to "0," no uid
is infusing. Be sure the access device is patent and has been ushed
per facility policy to maintain patency during this time period. If
a KVO rate is programmed, then that KVO rate will be used in the
Delay Start period.
1 0 5
e display then returns to the next eld in
e cursor will automatically scroll through
S c r e e n