Moog Curlin 6000 CMS IOD User Manual page 125

Ambulatory infusion system

Advertisement

I n t e r m i t t e n t
Note:
e top line of these running screens will display the name of the running protocol when
programmed in the Library mode, or it will display "Unidenti ed" when programmed outside of
the library.
e infusion can be paused at any time by pressing the PAUSE key.
blinks, and the Resume/Repeat Rx/New Program Menu screen is displayed.
e screen o ers the user three options:
Re sum ing t he Curre nt Cont a ine r
1. RESUME: To resume the therapy in progress from where it left o , press the YES/ENTER
key at this eld.
e pump will return to the Run Options Screen, and you can resume the
therapy.
Note: If you wish to hang a new container, move the cursor to "REPEAT Rx" (see following section).
a. Press the RUN key to resume the therapy.
b. Press the NO key to go back to the programming screen to make changes in the therapy
program (this feature may be blocked by a security lock, thus not allowing changes to the
program unless an access code is given). Also, remember that no changes are allowed in the
rst prescription menu if an infusion is in progress.
c. Press the OPTIONS or PRIME keys if either of these functions is desired at this time.
2. REPEAT Rx: To repeat the present therapy and add another IV bag or reservoir with the same
prescription, press the YES/ENTER key at this eld. If more than one therapy is
I n f o r m a t i o n
I nt e rrupt ing a n I nfusion
H a nging a N e w Cont a ine r
1 1 4
S c r e e n s
e yellow standby LED light