Damage to the pump due to dropping or impact while the door is open can cause intermittent
free ow conditions and loss of pump accuracy, but may not be visible through inspection.
Volumetric testing is the only method to con rm proper function.
NOTE: Anti-siphon Valve (ASV) administration sets can be used to provide additional
free- ow protection.
Do not subject the pump to dropping or hitting against a hard surface. If at any time the
pump is dropped or hit, the pump must be checked for volumetric accuracy prior to reuse.
pump can only be brought back into service if the volumetric accuracy test passes per the user
manual. If the pump fails volumetric accuracy it must be returned to Moog Medical Devices
Group for evaluation.
Note:
is procedure does not replace the recommended annual preventive maintenance.
e pump should receive preventive maintenance by a quali ed technician at least
annually.
•
Do not use this pump with a pressure cu applied to the IV bag or medication reservoir bag,
and do not unduly squeeze or compress the bag during a running infusion.
•
Remove all air from the administration set and IV bag or reservoir before connecting it to a
patient's access site.
•
Do NOT prime the administration set while it is connected to the patient's access device.
Doing so could result in overdosing the patient and could cause injury or death.
•
Do not attempt to open the pump's housing. All service problems should be referred to an
appropriate service technician.
•
Danger: Risk of explosion if used in the presence of ammable anesthetic or explosive gasses.