ITEM
Down Occlusion
Detection Time
Microbore tubing with
occlusion at distal Luer
Programmable Time (HH:MM)
Continuous
TPN
Intermittent
Variable
PCA
Bolus Volumes
Microbore tubing with
occlusion at distal Luer
SPECIFICATION
For non- ltered sets:
Rate
Low Pressure
0.1 mL/hr
≤ 1 hr. 30 minutes
1 mL/hr
≤ 7 minutes
10 mL/hr
≤ 40 seconds
50 mL/hr
≤ 6 seconds
125 mL/hr
≤ 3 seconds
For ltered sets:
Rate
Low Pressure
0.1 mL/hr
≤ 9 hr. 15 minutes
1 mL/hr
≤ 50 minutes
10 mL/hr
≤ 4 min. 30 seconds
50 mL/hr
≤ 50 seconds
125 mL/hr
≤ 20 seconds
00:01 to 288:00 HH:MM
00:01 to 288:00 HH:MM
00:01 to 288:00 HH:MM
00:01 to 288:00 HH:MM
No time limit
Low Pressure: <0.1 mL
High Pressure: <0.1 mL
1 6 6
High Pressure
≤ 2 hrs. 15 minutes
≤ 1min. 30 seconds
<45 seconds
≤ 15 seconds
≤ 5 seconds
High Pressure
≤ 20 hrs. 30 minutes
≤ 2 hrs. 5 minutes
≤ 10 min. 30 seconds
≤ 3 minutes
≤ 50 seconds