A i r - I n - L i n e
Also, when the Air-In-Line Sensor is completely disabled and the infusion goes to an "Infusion
Complete" or "Empty Bag" status, the Air Sensor setting will automatically change to 2 mL.
7. AUDIO (Audio Volume):
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting).
When you select a number, the pump will beep at that audio level. Press the NO key and enter
a new setting (1 – 9). Repeat until the desired setting is reached, and then press YES/ENTER to
select it and move the cursor to the next eld.
8. PWR CK? (Power Check):
It will display the active power source of the pump, whether internal batteries or external power.
If the active source is the batteries, a graph will indicate approximately how much charge remains
in the batteries. To access this information, select "ON," and press YES/ENTER.
Source Screen will display for ve seconds and return to the Options Menu.
9. ACCEPT OPT?: If the changes made in the Options Menu are acceptable, press the YES/
ENTER key. If the changes are not acceptable, use arrow keys to cursor up to any eld for
corrections, or abort all changes made in this menu by pressing the NO key and return to the
previous screen.
ere are two remaining selections to consider from the Run Options Screen: NO to REVIEW and
RUN to START.
Revie w ing or Cha nging t he Progra m
To review or make any changes in the prescription programming, press the NO/CHANGE key at
the Run Options Screen, and the prescription screens will reappear (changes may be inhibited by
lock setting). Changes to the prescription can be made before the start of the therapy; however,
once a therapy has started, no changes can be made to the Pre-Prescription Screen. It will appear
only for review purposes. Attempts to change the Pre-Prescription Menu during a new program
review will cause an alert screen to be displayed, notifying the programmer that changes will require
reprogramming of the prescription.
RUN: When all parameters are correct and the pump has been primed properly and connected
to the access site, press RUN at the Run Options Screen to start the therapy. If a delayed start is
programmed, the Delay Start Count Down Screen will appear. If no delay start is programmed or
the delay has elapsed, the pump will begin infusing, and the following screen displays the status of
the therapy:
S e n s o r
S e t
a t
2 m L
is feature allows you to adjust the audio volume of the pump's
is feature allows you to periodically recheck the battery charge.
St a r t ing t he I nfusion
9 9
A l e r t
S c r e e n
e Power