NO FOOD Alarm
If re-priming of the pump delivery set is required, it is advisable to disconnect
the enteral adapter from the patient's feeding tube while re-priming.
1. Is bag of pump
delivery set empty?
NO
2. Is air present in
the tubing?
NO
3. Is formula
blenderized or
aggressively mixed?
NO
4. Are air bubbles
trapped inside the
cassette?
ERRA, ERRD or ERRF Alarms
1. Was pump door
open while pump
was turned on or
while running?
LOAD SET Alarm
1. Is door closed
securely?
YES
2. Is door cracked
or tab on inside
of door broken?
EnteraLite® and Infinity ® are trademarks of Zevex, Inc.
MOOG® is a registered trademark of Moog, Inc.
© 2016 ZEVEX Inc. 26084-001 Rev. B
YES
Refill bag, re-prime delivery set and resume feeding.
Has air been removed from bag of pump delivery set?
YES
IF NO:
Manually prime tubing by
turning bag upside down, gently
squeezing lower half of bag while
pinching the teal colored tubing just
below the " "drop symbol.
Hold this position until all air is
removed from bag and tubing.
Let formula sit for 10-15 minutes before preparing pump delivery set
YES
for feeding. If foam is present in tubing, flush by priming pump
delivery set either manually or with the pump and resume feeding.
Remove tubing from pump and manually prime tubing by turning bag
upside down, gently squeezing lower half of bag while pinching the
YES
teal colored tubing just below the drop symbol. Hold this position until
all air has passed through the teal segment of the tubing.
NO
MOOG Customer Service, (800) 970-2337 for assistance.
Turn pump o . Unload and reload delivery set cassette.
YES/
Make sure door is closed securely, then turn pump back on.
UNSURE
If alarm continues, contact MOOG Customer Service,
NO
NO/
Turn pump o . Unload and reload delivery set cassette.
UNSURE
Make sure door is closed securely, then turn pump back on.
Contact MOOG Customer Service, (800) 970-2337
YES
Contact MOOG Customer Service, (800) 970-2337
NO
TROUBLESHOOTING
GUIDE
IF YES:
the PRIME key to move
the air in the tubing past
pump delivery set
cassette and resume
feeding.
If alarm continues, contact
(800) 970-2337 for assistance.
to order a new door.
for assistance.
Press and hold
Need help?
Do you have a question about the Enteralite Infinity and is the answer not in the manual?
Questions and answers