the "NEXT?" eld and press the YES/ENTER key to move on to the next screen.
At the following screens, the prescription information is entered to complete the programming of
the therapy.
Variable
erapies are programmed in individual dose screens, from one up to twenty-four. Before
information about each individual dose is entered, the bag volume and number of total doses per
bag are entered on the initial programming screen. Once the total number of doses is entered and
accepted, the pump will subsequently display a screen for each dose to program the speci c dose
amount, rate, and time:
V a r i a b l e
1. BAG Vol (IV Bag Volume): Use the numeric keys to enter the amount of uid in the IV Bag
or Medication Reservoir (include the over ll volume).
Note:
e volume used for priming via the pump or for KVO is deducted from the Bag
Volume.
erefore, Bag Volume should be equal to or greater than the Amount To Be
Infused to start the infusion.
When the correct amount has been entered, press the YES/ENTER key to accept this eld and
move on to the next.
2. # Doses:
e Variable
individual programs or "doses." Enter the total number of doses to be delivered from the
previously entered bag volume.
that will subsequently display. When the correct number of doses has been entered, press the
YES/ENTER key to accept this eld and move on to the next.
3. KVO RATE:
is eld requires data entry using the numeric keys to set the rate at which uid
will be delivered in a "Keep Open" state.
set from 0 to 10 mL/hour. All dose rates (mL/hr) must be greater than the KVO rate. When
the desired KVO rate has been entered, press the YES/ENTER key to accept and move the
cursor to the next eld.
4. NEXT?: If there are any changes to be made, use the up or down arrow keys to locate the eld
to be changed, use the numeric keys to enter the corrected information, and press YES/ENTER
to accept the new data. If all the elds are acceptable, move the cursor to the "NEXT?" eld and
press the YES/ENTER key to move on to the next screen.
P r e s c r i p t i o n
e range for this eld is 1 to 9999.
erapy can be programmed to deliver from one to twenty-four
is number dictates the number of dose programming screens
e pump defaults to a rate of 0.1 mL/hour, can be
1 2 1
M e n u 1
Need help?
Do you have a question about the Curlin 6000 CMS IOD and is the answer not in the manual?
Questions and answers