LCD Display:
"***ALERT *** CALCULATED AMT OUT OF RANGE"
Alert:
Invalid calculated amount for infusion
Cause:
e calculated amount for the infusion, based on the values entered for rate
and time, exceeds either the pump limit or the value entered for bag volume.
Resolution:
Recheck data entry for Rx.
LCD Display:
"***ALERT *** CALCULATED RATE OUT OF RANGE"
Alert:
Invalid calculated rate of infusion
Cause:
e calculated rate for the infusion, based on the values entered for Volume/
Amount and Time, exceeds either the minimum or maximum rate limit of the
pump or the maximum rate programmed in MedLMTS
Resolution:
Recheck data entry for Rx.
LCD Display:
"***ALERT *** CALCULATED TIME OUT OF RANGE"
Alert:
Invalid calculated time
Cause:
e calculated time for the therapy, based on the values entered for Volume/
Amount and Rate, exceeds the therapy time limit of 288:00 HHMM (167:59
HHMM Software Versions below R37C).
Resolution:
Recheck data entry.
range.
LCD Display:
"***ALERT *** Changes Require Re-programming. YES to Continue NO to
Return"
Alert:
Changes to the data will require re-entry of prescription.
Cause:
Attempted to change the Pre-Prescription Screen prior to start of infusion.
Resolution:
If a change to the programmed prescription is required, press YES.
Note: Selecting YES will clear the entire prescription and require re-entry the
If no change is required, press NO to preserve the current prescription and
return back to the Pre-Prescription screen.
LCD Display:
"***ALERT *** DATE/TIME CHANGE NOT ALLOWED"
Alert:
Not allowed to change date or time.
Cause:
Attempt made to modify system clock while a therapy/delay is in progress.
Resolution:
If the therapy in progress is to be continued, wait until the infusion is complete
and then change the date and time.
If it is not necessary to preserve the therapy in progress or any other
programmed therapy, use the "CLEAR Rx" feature of the Clinician
e time calculated by the pump cannot exceed the pump
prescription.
1 3 6
limits.