If no air-eliminating lter is being used, press NO.
resume and the Air-In-Line sensitivity will be set to 2 mL for the remainder of
the infusion.
Note: If the screen is not acknowledged with a YES or NO key press, the Air-
LCD Display:
"***ALERT *** PUMP WILL SENSE 2 ML OF AIR"
Alert:
Air-In-Line sensitivity will be set to 2 mL.
Cause:
NO was pressed at the "AIR SENS OFF. Using In Line Filter?" alert screen.
e air in line sensitivity will be set to trigger an AIR IN LINE alarm if 2 mL of
air is detected.
Resolution:
If an air eliminating lter is being used, press YES when the "AIR SENS OFF.
Using In Line Filter?" screen is displayed.
indicated under "AIR SENS OFF. Using In Line Filter?" alert explanation.
LCD Display:
"***ALERT *** AMT NOT 0 OR CURRENT DOSE AMT"
Alert:
Unable to change clinician dose while clinician dose in progress.
Cause:
A clinician dose is in progress, and a new clinician dose value is entered that is
not "0" or the current dose amount.
Resolution:
If the current clinician dose is to be continued, change the "Clin DOSE" value
back to the original value and accept the eld and menu. If the current clinician
dose is to be terminated, which will allow a new clinician dose to be
programmed, enter a value of 0 in the "Clin DOSE" eld and accept the eld
and menu.
DOSE" value can be entered.
LCD Display:
"***ALERT *** AMT TBI GREATER THAN BAG VOL"
Alert:
Value entered is greater than value in Bag Vol eld.
Cause:
Occurs when the amount entered for the "Amt TBI" or "Vol TBI" eld is
greater than the value in the bag volume eld.
Resolution:
Recheck the data entry.
LCD Display:
"***ALERT*** AMT TBI GREATER THAN BAG VOL. YES to Continue
NO to Return"
Alert:
Insu cient volume
Cause:
Occurs when bag volume is less than amount to be infused. Can occur when
pump's priming feature is used as bag volume is depleted by the amount used in
priming.
Resolution:
To modify the Bag Volume or Amt/Vol TBI, press NO to return to the Run
Options Screen.
NO to REVIEW on the Run Options Screen.
In-Line sensitivity will be set to 2 mL when the screen is cleared (i.e.
alarm occurs while screen is displayed).
e next time that the Options Menu is accessed a new "Clin
e programming screens will then be accessible by pressing
is will disable air detection as
1 3 4
e infusion will start/