A i r - I n - L i n e
is message alerts you to the Air-In-Line Sensor being set to "OFF."
answering yes or no to whether the administration set being used has an air-eliminating lter.
If you press the YES key to indicate that there is an air-eliminating lter in use, the Air-In-Line
sensor will be completely disabled and the pump will start the infusion.
If you press the NO key to indicate a lter is not in use, the Air-In-Line Sensor will be set to
detect the presence of any contiguous 2-milliliter air in the line.
then brie y appear on the screen. If the alert occurred at the Run Options Screen, the pump
will start the infusion.
A i r - I n - L i n e
Also, when the Air-In-Line Sensor is completely disabled and the infusion goes to an "Infusion
Complete" or "Empty Bag" status, the Air Sensor setting will automatically change to 2 mL.
9. AUDIO (Audio Volume):
alarm setting. You can enter a number from 9 (the loudest setting) to 1 (the quietest setting).
When you select a number, the pump will beep at that audio level. Press the NO key and enter
a new setting (1 – 9). Repeat until the desired setting is reached, and then press YES/ENTERto
select and move the cursor to the next eld.
10. PWR CK? (Power Check):
It will display the active power source of the pump, whether internal batteries or external power.
If the active source is the batteries, a graph will indicate approximately how much charge
remains in the batteries. To access this information, select "ON" and press YES/ENTER.
Power Source Screen will display for ve seconds and return to the Options Menu.
11. ACCEPT OPT?: If the changes made in the Options Menu are acceptable, press the YES/
ENTER key. If the changes are not acceptable, cursor up to any eld for corrections or abort all
changes made in this menu by pressing the NO key, and return to the previous screen.
ere are two remaining selections to consider from the Run Options Screen, NO to REVIEW and
RUN to START.
To review or make any changes in the prescription programming, press the NO/CHANGE key at
the Run Options Screen, and the prescription screens will reappear (changes may be inhibited by
lock setting). Changes to the prescription can be made before the start of the therapy; however, once
a therapy has started, no changes can be made to the Pre-Prescription
S e n s o r
S e n s o r
is feature allows you to adjust the audio volume of the pump's
is feature allows you to periodically recheck the battery charge.
Revie w ing or Cha nging t he Progra m
O f f
M e s s a g e
a t
2 m L
A l e r t
8 4
S c r e e n
e message will require
e alert message below will
S c r e e n