1. LOCK: When the lock is set to "OFF, " the clinician has access to all of the pump functions.
After the pump is completely programmed, change the lock to meet the security needs of this
therapy.
ere are four choices: "OFF," "1," "2," or "3." Each level higher limits the available
options. To change the lock setting, press the NO/CHANGE key until the correct lock setting
appears.
en, press YES/ENTER key to accept the new lock selection. Once the pump has been
set into a locked condition, any subsequent request to change the lock will require a clinician
access code. (See "Clinical Information and Access Code" chapter for access code information,
lock setting table and available features for each lock setting.)
2. HRLY TOT (Hourly Totals): To view hourly infusion totals press the NO/CHANGE key to
change OFF to ON and then press YES/ENTER.
e screen displays hourly infusion delivery for each hour (00-23) of the last 24 hours of
therapy in the unit selected.
a.
e highlighted eld has a marker
b. Use the up arrow key to scroll back through the last 24 hours.
c. If the infusion was not running, a series of ------ will appear for that hour.
d. To exit this screen, press the NO/CHANGE key.
3. CLR HRLY (Clear Hourly Totals): To clear Hourly Totals press the NO/CHANGE key to
change OFF to ON and then press YES/ENTER.
a. To clear the Hourly Totals, press the YES/ENTER key. All Hourly Totals will be cleared
and the cursor will return to the Options Menu.
b. To maintain the Hourly Totals, press the NO/CHANGE key. Hourly Totals will be
maintained and the cursor will return to the Options Menu.
c. A lock level of 2 or 3 will require an access code entry before allowing Hourly Totals
reset.
H o u r l y
T o t a l s
indicating the current hour in progress.
C l e a r
H o u r l y
1 2 6
e following screen will appear:
S c r e e n
e following screen will appear:
T o t a l s
S c r e e n