e. Air-In-Line Sensor - Reinstall the administration set properly and resume the infusion. From
the Options Menu set the AIL SENS (AIR IN LINE sensitivity) to 0.1mL. With the pump
running introduce an air bolus of approximately 0.5mL into the uid path at the input side
of the pump. Once approximately 0.1mL of air has past the AIL detector, an "ALARM AIR
IN LINE" will occur. PAUSE and PRIME the air bolus past the AIL sensor, then resume the
infusion.
f. Auxiliary Alarm - With the pump running observe the "Vol INF" (volume infused on the run
screen) and disconnect the pump from all power sources (unplug the AC Adaptor eliminator
and remove one C-cell battery). Observe that the Auxiliary Alarm System is activated, with
the red LED and beeper on constant (not intermittently beeping). With the battery still out
of the pump, press the ON/OFF button to turn o the auxiliary alarm. Replace the battery
in the battery compartment and turn on the pump. Resume the therapy and con rm that the
infusion continues where it stopped due to Auxiliary Alarm occurrence.
5. SET BIOMED CLINICIAN MENU ITEMS:
a. Turn the pump o and back on. While pump is performing Self Tests con rm that the
pump is set to the correct date and time. If date and time are not correct, program the
correct date and time from the BioMed/Clinician Menu.
b. Enter the BioMed/Clinician Menu using the Clinician Code.
c. Set the Lock Level to OFF in the CHANGE LOCK Menu.
d. Clear the Pt DATA from the pump.
e. Observe the pump maintenance date, top of the BioMed/Clinician Menu, to ensure
adequate time for the next use by patient.
f. Sign, date and return the check list to your record le. Pumps that pass can be used for
new patients. Pumps that fail are to be returned for service.