ree tamper-resistant lock settings to assist in maintaining patient compliance and safety.
Patient and therapy speci c programming features via the Moog Curlin Infusion BIOMED
SETUP and Options Menus (when not limited by the institution).
Ability, at the institution level, to con gure the pump for speci c administration route or units
of measure:
º
Units (Factory, mL, mg, mcg, mL xed, mg xed or mcg xed);
º
Route (Intravenous, Epidural, Subcutaneous, Intravenous
Subcutaneous xed).
Retention of programmed infusion settings until cleared by the clinician, thus eliminating the
need to reprogram before each use.
Mandatory requirement of two independent key press actions before a running pump can be
turned o , thus minimizing the accidental interruption of a therapy in progress.
Powered by two readily available, cost-e ective C-cell alkaline batteries.
Attractive yet robust design that is impact resistant and water resistant.
Audio alarms that can be adjusted from 1 (very quiet) to 9 (loudest) to meet speci c patient
and clinical setting needs but that cannot be totally disabled.
Four sizes of convenient carry packs to assist the ambulatory patient in maintaining
independence of lifestyle. Backpack will accommodate pump, AC Adaptor, and up to four-
liter IV bag.
Retained memory features that allow the pump to easily resume a therapy exactly from where
it left o when it is interrupted before completion or to repeat a therapy using the same
programmed prescription.
Compact, lockable safety shells that provide additional security when needed and that are
designed to be free standing, pole mounted, or placed in the convenient, soft, carry case.
Backlit illumination for display screen. Backlight remains on continuously if AC Adaptor is
in use.
Pump will not permit the user to enter any programming value that is outside of its
predetermined range.
Integral Up as well as Down Occlusion alarms to notify user if an occlusion is present on either
side of the pumping chamber.
Integral High Upstream pressure alarm to notify user of excessive upstream pressure.
Integral Air-In-Line alarm to notify user of presence of air in the tubing.
Adjustable Down Occlusion settings of "High" and "Low" to customize the pump's down
pressure requirements.
Patient side (downstream) "Line Pressure" displayed.