1. Press and release the PRIME key at the Run Options Screen.
2.
e following screen appears:
3. Disconnect the administration set from the patient's access site using aseptic protocols.
I M PORTAN T
I N FORM AT I ON
4. Release any clamps on the administration set.
5. Press and hold the PRIME key. Fluid will be pumped at a rate of 300 mL/hr through the
administration set as long as the PRIME key is pressed or until 6 mL is pumped.
will then beep once and require repressing and holding the PRIME key again until 6 more
milliliters is delivered. Continue this process until all air has been purged from the IV bag and
the administration set.
Note: If the bag volume is depleted during priming, a subsequent priming event cannot be
initiated.
While the PRIME key is depressed, the following screen appears:
e amount used for the prime session is accumulated and displayed on the bottom eld of this
screen.
6. When priming is complete, release the PRIME key.
7. To continue priming, press the PRIME key again. If priming is complete, press YES/
ENTER to exit and return to the Run Option Screen.
priming session displays on the bottom eld.
P r i m e
D i r e c t i o n
Priming with the set connected to the patient could result in overdose
and may cause injury or even death to the patient.
7 8
S c r e e n
e following screen will display:
e total amount of solution used in this