Error
LCD Displayed
Code
Message
162
Stuck 0 Key
Turn Pump OFF/ON or Call Provider
163
Stuck 1 Key
Turn Pump OFF/ON or Call Provider
Stuck 2 Key
164
Turn Pump OFF/ON or Call Provider
165
Stuck 3 Key
Turn Pump OFF/ON or Call Provider
166
Stuck 4 Key
Turn Pump OFF/ON or Call Provider
167
Stuck 5 Key
Turn Pump OFF/ON or Call Provider
168
Stuck 6 Key
Turn Pump OFF/ON or Call Provider
169
Stuck 7 Key
Turn Pump OFF/ON or Call Provider
170
Stuck 8 Key
Turn Pump OFF/ON or Call Provider
171
Stuck 9 Key
Turn Pump OFF/ON or Call Provider
Stuck UP Key
172
Turn Pump OFF/ON or Call Provider
173
Stuck NO Key
Turn pump OFF/ON or Call Provider
174
Stuck YES Key
Turn Pump OFF/ON or Call Provider
175
Stuck DOWN Key
Turn Pump OFF/ON or Call Provider
176
Stuck RUN Key
Turn Pump OFF/ON or Call Provider
177
Stuck HELP Key
Turn Pump OFF/ON or Call Provider
178
Stuck PRIME Key
Turn Pump OFF/ON or Call Provider
182
Turn Pump OFF/ON
Call Provider
200
Turn Pump OFF/ON
Call Provider
201
Turn Pump OFF/ON
Call Provider
202
Turn Pump OFF/ON
Call Provider
Turn Pump OFF/ON
203
Call Provider
Turn Pump OFF/ON
204
Call Provider
Turn Pump OFF/ON
205
Call Provider
Corrective Action
Probable Cause
Stuck 0 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
Stuck 1 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck 2 Key
Otherwise Report Error Code To Factory
Stuck 3 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck 4 Key
Otherwise Report Error Code To Factory
Stuck 5 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
Stuck 6 Key
Otherwise Report Error Code To Factory
Stuck 7 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
Stuck 8 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
Stuck 9 Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck Up Arrow Key
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck No / Change Key
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck Yes / Enter Key
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck Down Arrow Key
Otherwise Report Error Code To Factory
Stuck Run / Pause Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
Stuck Help / Options Key
If Key Was Pressed, Release It
Otherwise Report Error Code To Factory
If Key Was Pressed, Release It
Stuck Prime / Bolus Key
Otherwise Report Error Code To Factory
Invalid Key
Report Error Code To Factory
Invalid System State
Report Error Code To Factory
Bad Power Up Test Number
Report Error Code To Factory
Bad Power Up State
Report Error Code To Factory
Bad A to D Max Address
Report Error Code To Factory
Report Error Code To Factory
Bad Run State
Bad Pump Status
Report Error Code To Factory
1 5 1