LCD Display:
"***ALERT *** VALUE
Alert:
Value entered is below the Library Limit
Cause:
e value entered is out of range of the soft limit setting(s) for a downloaded
therapy.
Resolution:
e screen will ask the user to override the soft limit value. Pressing YES will
accept the new value.
LCD Display:
"***ALERT *** EMPTY LIBRARY"
Alert:
Empty Library
Cause:
Occurs when "LIBRARY" is selected and no library is present.
Resolution:
Load a Library or program the pump using the "PROGRAM" option.
LCD Display:
"***ALERT *** PUMP DISABLED CALL PROVIDER"
Alert:
e pump is disabled.
Cause:
e pump calibration data is invalid.
Resolution:
e pump needs to be returned for service.
LCD Display:
"***ALERT*** DOOR OPEN or SET NOT PROPERLY INSTALLED"
Alert:
Door is opened or administration set is not properly installed while the pump is
running.
Cause:
Door is opened or administration set is not properly installed while the pump is
running
Resolution:
Check the door latch assembly and securely close the door. If the door was
closed properly, re-install the set according to the installation instructions.
See Figure 1.8 for instructions on closing door.
Alarms are conditions that require correction by the user and may be a safety issue but are not
system errors of the pump. All alarm conditions stop the infusion.
is at the "Infusion Complete" alarm, which allows the KVO rate, when programmed, to continue
infusing. When an alarm occurs, a message is displayed on the screen, the audio alarm is sounded,
and the red LED light ashes.
SILENCE key can be pressed to disable the audio alarm for one-minute intervals.
alarm must be resolved, or the alarm will recur when the infusion is restarted. To exit an alarm
screen, press the PAUSE key, resolve the alarm condition, and then resume the therapy by selecting
"RESUME" from the Pause Menu.
e following is a selected and alphabetical list of the alarms, conditions that cause the alarms,
messages displayed on the screen, and actions necessary to resolve them:
LCD Display:
"***ALARM*** AIR IN LINE"
LIMIT"
ALARM S
e message stays on the screen until the pause key is pressed.
1 4 5
e only exception to this rule
e cause of the
e