c. BOLS INT (Bolus Interval): Use the numeric keys to enter the new bolus time interval and,
when correct, press YES/ENTER to accept and move to next eld.
is 0 to 60.
d. #BOLS /hr (number of boluses per hour): Use the numeric keys to enter the new number
of boluses per hour and, when correct, press YES/ENTER to move to the nal step in the
MedLMTS™ menu.
e. ACCEPT? YES/NO: If necessary, use the arrow keys to move to any eld and change the
data entered. To abort this screen entirely, press NO at this prompt. Any data entered on this
display will be aborted and the display will return to the options menu. If the parameters
entered are correct, press the YES/ENTER key, and the new rate or rates will be accepted.
e display returns to the Options Menu when this screen is exited.
f. New values will be in e ect when the Options Menu is accepted.
7. DN Occlu (Down Occlusion):
the line between the pump and the access site, e.g., a kink in the tubing between the pump
and the patient, a blockage in the access site, or a clamp left on the distal tubing.
options for this eld: the "Low" setting triggers the alarm when it detects in-line pressures at 400
mmHg ±150 (8 ±3 psi "pounds per square inch"), or the "High" setting (default setting) alarms
when it detects in-line pressures at 900 mmHg ±150 (18 ±3 psi). Press the NO/CHANGE key
to change the selection. When the correct pressure setting appears, press the YES/ENTER key to
accept it and move the cursor to the next eld. If setting is changed, the new settings will remain
in e ect until a new therapy is programmed.
8. AIR SENS (Air In Line Sensor):
will detect in the administration set tubing before it goes into an alarm state.
"0.1 mL," "0.5 mL" (the default setting), "2 mL," and "OFF." Press the NO/CHANGE key to
change the options until the desired setting appears on the display, and then press YES/ENTER
to select and move the cursor to the next line. If setting is changed, the new settings will remain
in e ect until a new therapy is programmed.
Whenever the Air-In-Line Sensor is set to OFF, Moog recommends
I M PORTAN T
the use of an air-eliminating lter.
I N FORM AT I ON
When the Air-In-Line Sensor is set to "OFF" and the RUN key is pressed at the Run Option
Screen, the alert message below will appear on the screen. It will also appear if the Air-In-Line
is changed to "OFF" from the OPTIONS menu when an infusion is in progress and the AIR
SENS eld is accepted.
e range for this eld is 0 to 15.
is feature allows the pump to sense an occlusion pressure in
is feature allows you to select the amount of air the pump
8 3
e range for this eld
ere are two
e choices are