milliliters is delivered. Continue this process until all air has been purged from the IV bag and
the administration set.
Note: If the bag volume is depleted during priming, a subsequent priming event cannot be
initiated.
While the PRIME key is depressed, the following screen appears:
e amount used for the prime session is accumulated and displayed on the bottom eld of this
screen.
6. When priming is complete, release the PRIME key.
7. To continue priming, press the PRIME key again. If priming is complete, press YES/ENTER to
exit and return to the Run Options Screen.
session displays on the bottom eld.
Note:
e amount of uid used in the priming process will be deducted from the Bag Volume
Total but will NOT be counted in the amount infused because this amount is not to be
delivered to the patient.
TBI before you can proceed with the infusion. If it is not, an alert will appear stating, "AMT
TBI GREATER THAN BAG VOL." You have the option to continue by pressing the YES/
ENTER key or return to the Run Options Screen by pressing the NO key.
Once the prescription is programmed, there are more customizing features available for each therapy.
Press the OPTIONS key from the Run Options Screen to access these options.
menu can also be accessed while the pump is running.
Note: After the infusion has started, any changes made through the OPTIONS menu involving:
"HRLY TOT" and
"CLR HRLY"
will return you directly to the "ACCEPT OPT?" eld.
e following screen displays the Intermittent
menu, use the up and down arrows to move the highlight bar to the desired option. Lock settings
may require an access code before the option can be changed. Any new setting, except "CLEAR
HOURLY" screen acceptance becomes e ective when the Options Screen is exited by pressing
e total amount of solution used in this priming
e bag volume should still be equal to or greater than the Amount
Opt ions
erapy Options available. When you open this
1 0 9
e following screen will display:
e OPTIONS

Advertisement

loading
Need help?

Need help?

Do you have a question about the Curlin 6000 CMS IOD and is the answer not in the manual?

Questions and answers

This manual is also suitable for:

Curlin 6000 cms6000 cms6000 cms iod