Download Print this page

Alarms; Power Loss Alarm; Sash Alarm; Blower Failure Alarm - Labconco Logic Vue User Manual

Class ii enclosure

Advertisement

Alarms

Any alarm that activates requires the user's immediate attention, and some form of
actionable response to clear the alarm.

Power Loss Alarm

The enclosure has lost power. See Figure 6-19.
Press
[OK]
on the keypad to acknowledge that
a power loss occurred.
Note: This alarm will activate any time power is
cycled, including turning the System Reset
Switch (see Figure 9-2) off and back on.

Sash Alarm

Both Sashes or one Hatch is not at the proper
operating position. Return the Hatch or one Sash
to proper working height.

Blower Failure Alarm

A blower motor has failed, or the motor and
display circuit board are not communicating
properly. Press
[BLOWER]
clear the alarm.
DO NOT USE THE ENCLOSURE UNTIL THE PROBLEM HAS BEEN CORRECTED.
on the keypad to
81
Figure 6-19
Figure 6-20
Figure 6-21
BLOWER FAILURE
DETECTED
CONTACT
CERTIFIER

Advertisement

loading
Need help?

Need help?

Do you have a question about the Logic Vue and is the answer not in the manual?