Labconco Logic Vue User Manual page 151

Class ii enclosure

Advertisement

7. Do not change the
[OK/MUTE]
8. The
Sash Height
[OK/MUTE]. You will return to the Configuration / Calibration submenu within the
password protected area.
Once the Airflow Sensor is configured, it can now be calibrated, follow this procedure:
If you left the password-protected screens, re-enter the correct password:
[UP] [TIMER] [DOWN]
PASSWORD
* * * * *
Note: The sashes must not be closed when pressing [OK/MUTE]. Access will not be
granted if both Sashes are closed.
Proceed to the following page.
Canopy
selection, press [OK/MUTE].
SASH HEIGHT
8 INCH
selection will display next, and should NOT be changed. Press
[OK/MUTE]. Select Calibration, press [OK/MUTE].
Password
Entered
151
[LIGHT]
CONFIGURATION
CALIBRATION