Working in the Logic Vue
Working in the Logic Vue enclosure requires preparation, planning, and an
understanding of aseptic technique to work safely. The steps below represent the
minimum requirements before, during, and after working in the enclosure. Consult your
EH&S team for additional SOP requirements.
Planning
1. Thoroughly understand procedures and equipment required before beginning
work.
2. Arrange for minimal disruptions, such as room traffic or entry into the room while
the cabinet is in use.
Equipment Placement
When placing large equipment inside the Logic Vue enclosure, make certain it is
centered in the work area (as close as reasonably possible). Remove any safety covers
to allow HEPA-filtered air to flow through the equipment. If the covers are not removed,
high velocity regions can develop which impairs product and/or personnel protection.
See Figure 9-5. For additional details, see
High Velocity Regions
(if equipment cover
panels are not removed)
Equipment Placement
Figure 9-5
103
in Section 4.
Need help?
Do you have a question about the Logic Vue and is the answer not in the manual?