The HEPA filters in the enclosure will gradually accumulate airborne particulate
matter from the room and from work performed in the cabinet. The rate of
accumulation will depend upon the cleanliness of the room air, operating time
and the nature of work being done in the enclosure. The Filter Gauge accurately
displays the amount of filter life remaining.
Proper operation of the enclosure depends largely upon its location and the
operator's work habits. Consult
Logic Vue
for further details.
When surface disinfecting the enclosure:
o Avoid splashing the disinfecting solution on skin or clothing.
o Ensure adequate ventilation.
o Carefully follow the disinfectant's safety instructions.
o Always dispose of disinfecting solutions in accordance with local and national
laws.
o DO NOT allow disinfectants with high concentrations of free chlorine to
contact the stainless steel components of the enclosure for a long period of
time. Free chlorine will corrode stainless steel after extended contact.
The enclosure should be decontaminated for any of the following reasons:
o Before maintenance work requiring entry into contaminated areas.
o Before contaminated HEPA filter changes.
o Before performing certification tests requiring entry into contaminated areas.
o Before relocating a contaminated enclosure.
o Before changing research programs that used hazardous materials.
o After the gross spill of biohazardous material or toxic chemicals.
Never work with biohazards or hazardous particulate through the rear sash.
Opening the rear sash will jeopardize product protection and may jeopardize
personnel protection. The rear sash opening and rear hatch must only be
opened for maintenance or service access to the equipment inside.
This product must be powered with a dedicated electrical circuit. The dedicated
supply electrical circuit must be wired through an appropriately rated circuit
breaker suitably located within 10 feet (3m) of the product and easily reachable.
The switch must be marked as the disconnecting device for the product.
Le circuit électrique d'alimentation dédié doit être câblé via un interrupteur de
calibre approprié situé à moins de 3 m (10 pieds) du produit et facilement
accessible. L'interrupteur doit être marqué comme le dispositif de déconnexion
du produit.