Cabinet Start-up

1. Open the Front Sash.
2. Turn on the light and enclosure blowers.
3. Check the air slots around the perimeter of the work surface for obstructions.
4. Allow the enclosure to operate until the Home Screen is shown (Figure 9-6).
BLOWER
STARTING
BLOWER
STARTING
12:39
5. Wash hands and arms thoroughly with germicidal soap.
6. Wear appropriate personnel protective equipment (PPE).

Wipe Down

1. Raise the Front Hatch to its full open position. Mute the alarm by pressing
[OK/MUTE].
2. Wipe down the interior surfaces of the cabinet with 70% ethanol, or a suitable
disinfectant, and allowed to dry.

Loading Materials and Equipment

1. Only load the materials required for the procedure. Do not overload the
enclosure.
2. Do not obstruct the front, side, or rear return air slots.
3. Large objects should not be placed close together.
4. Close Front Hatch.
5. After loading the enclosure, wait two to three minutes to purge airborne
contaminants from the work area.

Work Techniques

• Keep all materials at least 4 inches (100 mm) inside from the sash, and perform
all contaminated operations as close to the center of the work area as possible.
• Segregate clean materials from contaminated materials in the work area.
• Arrange items to minimize movement of contaminated materials into clean areas.
• Keep all discarded contaminated material to the rear of the work area.