General Console Requirements

2-2-1

Console Environmental Requirements

Table 2-2
Operating
Temperature
10 to 40
(50 to 104
If the system has been in storage or has been transported, please see the acclimation requirements
CAUTION
before powering
!! CAUTION:
2-2-1-1

Cooling

The cooling requirement for the Voluson® 730Expert is 3446 BTU/hr. This figure does not include
cooling needed for lights, people, or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
2-2-1-2

Lighting

Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2-2

Electrical Requirements

NOTE:
GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound
equipment. This dedicated power shall originate at the last distribution panel before the system.
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the Ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2
3
DRAFT (A
Environmental Requirements
Operating
Humidity
o
C
30 to 80% rH
o
non-condensing
F)
and/or using the system (see:
ON
Section 2-2 - General Console Requirements
29, 2008)
PRIL
Heat Dissipation
3446 BTU pr hour
Section 3-2-2 "Installation Warnings" on page
V
® 730E
OLUSON
XPERT
S
ERVICE
Storage
Temperature
Storage Humidity
o
-10 to 40
C
< 90% rH non-
o
condensing
(14 to 104
F)