General Console Requirements; Environmental Requirements; Cooling; Lighting - GE Voluson P8 Basic Service Manual

Hide thumbs Also See for Voluson P8:
Table of Contents

Advertisement

GE
RAFT
DIRECTION 5775469, R
Section 2-2

General Console Requirements

2-2-1

Environmental Requirements

Table 2-2
Operating
Temperature
18 to 30
(64 to 86
If the system has been in storage or has been transported, please see the acclimation requirements
CAUTION
before powering
!! CAUTION:
2-2-1-1

Cooling

The cooling requirement for the Voluson™ P8/Voluson™P6 is 1200BTU/hr. This figure does not include
cooling needed for lights, people, or other equipment in the room.
NOTE:
Each person in the room places an additional 300 BTU/hr. demand on the cooling system.
2-2-1-2

Lighting

Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2-2

Electrical Requirements

NOTE:
Voluson™ P8/Voluson™P6 requires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the
system.
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size Ground wire from the distribution panel to the Ultrasound outlet.
NOTE:
Please note that image artifacts can occur, if at any time within the facility, the Ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2
3
DRAFT (J
EVISION
Environmental Requirements
Operating
Humidity
o
C
30 to 80% rH
non-condensing
o
F)
and/or using the system (see:
ON
Section 2-2 - General Console Requirements
19, 2018)
ULY
Heat Dissipation
Temperature
1200BTU/h
without on-board
peripherals
Section 3-2-2 "Installation Warnings" on page
V
™ P8/V
OLUSON
OLUSON
B
S
ASIC
ERVICE
Storage
Storage Humidity
o
-10 to 50
C
10 to 90% rH
non-condensing
o
(14 to 122
F)
™P6
M
ANUAL
3-3).

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Voluson p6Voluson p6 bt1857782185778216

Table of Contents