GE LOGIQ Totus Basic Service Manual page 637

Table of Contents

Advertisement

Direction 5936427, Rev. 1
Type BF Patient Leakage Current Test - Probes (This test should be done for all probe)
Table 10-21
Tester
Neutral
Closed
Closed
Closed
Closed
Open
Open
Type BF Patient Leakage Current Test - Mains to Probe (This test should be done for all probes)
Table 10-22
Tester
Neutral
Closed
Closed
Final Check. All system covers are in place. System scans with all probes as expected.
Accepted by: ______________________________________________________________________
10-7-6
Ultrasound Equipment Quality Check (EAC and IQC)
Download and use the latest version of these forms. They can be retrieved from MyWorkshop.
EQC -- Refer to DOC0929340 in MyWorkshop.
GE HealthCare CONFIDENTIAL
Tester Polarity
Ground
Switch
Switch
Normal
Closed
Normal
Open
Reversed
Closed
Reversed
Open
Normal
Closed
Reversed
Closed
Tester Polarity
Ground
Switch
Switch
Normal
Closed
Reversed
Closed
Chapter 10 Care & Maintenance
LOGIQ Totus Basic Service Manual
Max Value
Value
Allowed
Measured
0.1mA
0.5mA
0.1mA
0.5mA
0.5mA
0.5mA
Max Value
Value
Allowed
Measured
5mA
5mA
OK?
Comments
OK?
Comments
10 - 35