Section 2-2
General Console Requirements
2-2-1
Console Environmental Requirements
Table 2-2
Operating
Temperature
10 to 40
(50 to 104
2-2-1-1
Cooling
The cooling requirement for the Voluson® 730 is 2500 BTU/hr This figure does not include cooling
needed for lights, people, or other equipment in the room. Each person in the room places an additional
300 BTU/hr. demand on the cooling system.
2-2-1-2
Lighting
Bright light is needed for system installation, updates and repairs. However, operator and patient
comfort may be optimized if the room light is subdued and indirect. Therefore a combination lighting
system (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source
of EMI which could degrade image quality. These controls should be selected to minimize possible
interference.
2-2-2
Electrical Requirements
NOTE:
GE Medical Systemsrequires a dedicated power and ground for the proper operation of its
Ultrasound equipment. This dedicated power shall originate at the last distribution panel before
the system. The dedicated line shall consist of one phase, a neutral (not shared with any other
circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet. Please
note that image artifacts can occur, if at any time within the facility, the ground from the main
facility's incoming power source to the Ultrasound unit is only a conduit.
2-2-2-1
Voluson® 730 Power Requirements
Table 2-3
2-2-2-2
Inrush Current
Inrush current is not a factor to consider due to the inrush current limiting properties of the power
supplies.
2-2
U
RETZTECHNIK
LTRASOUND
1
Environmental Requirements
Operating
Humidity
o
C
30 to 80% rH
o
non-condensing
F)
Electrical Specifications for Voluson® 730
Voltage
Tolerances
100 VAC
115 VAC
130 VAC
230 VAC
240 VAC
Section 2-2 - General Console Requirements
Heat Dissipation
2500 BTU pr hour
Current
±10%
10.10 A
±10%
8.80 A
±10%
7.80 A
±10%
4.40 A
±10%
4.20 A
V
® 730 S
OLUSON
ERVICE
Storage
Temperature
Storage Humidity
o
-10 to 40
C
< 90% rH non-
o
condensing
(14 to 104
F)
Frequency
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
50, 60 Hz (±2%)
M