GE LOGIQ Totus Basic Service Manual page 172

Table of Contents

Advertisement

Direction 5936427, Rev. 1
Table 4-2 General System Functional Checks (Continued)
Section
4-3-22
4-3-25
4-3-24
4 - 6
GE HealthCare CONFIDENTIAL
Functional Check
VSIO (ECG)
ECG Checks
Active ECG:
Utility > Applications > Settings > ECG >
Show ECG Tools.
Connect cables. Press Display ECG Tab on
TP. Verify ECG signal displayed on monitor.
Plug in the ECG cables with nothing attached
so they pick up noise.
Diagnostics
Battery Pack/
Battery Pack functional checks
VNAV
Volume Navigation with inside probe functional
checks
Section 4-2 - General Procedures
LOGIQ Totus Basic Service Manual
Debrief Script
Leakage Current measured at (record the value) and meets
allowable limits. Equipment passed all required checks and is
ready for use.