LOGIQ Totus Basic Service Manual
6-2-2

Access to monitor adjustments

All monitor settings have been optimized at the factory, so there should be no need for any further
adjustments. The default settings on Scan Screen are:.
Table 6-1
Room Profile Table
HDU Display
LCD Monitor
Room Profile: Dark
Room Profile: Dark
Brightness/Contrast: 52/50
Brightness/Contrast: 50/85
For more information the appropriate version of the LOGIQ Totus Basic User Manual Chapter 3, or the
appropriate LOGIQ Totus Release Notes.
1-4-1 "Label Location" on page
1-13.
To access the Scan Screen, navigate to Utility (second page) > Scan Screen
Figure 6-1 Scan Screen
Use the encoder knobs to adjust the Scan Screen settings (shown on the Scan Screen directly above
each encoder knob).
Figure 6-2 Operator Panel Encoder Knobs
Chapter 6 Service Adjustments
6 - 3