5936427, R
IRECTION
EV
8-8-17 Replacement of the Top Cover Assy-T
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5 /8-2-7 "PPE Required During Service" on page 8-6
Table 8-79
Manpower
One person
8-8-17-1
Preparations
1.) Remove the T2 Side Cover R Assy as described in
Assy" on page 8-136
8-8-17-2
Removal Procedure
1.
Lift the elevation arm to its highest position
Pull the Top Cover Assy-T to remove from the
2.
console.
8-8-17-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform "Functional Checks after FRU replacement (Debrief Guideline)" as described in
12 on page 8-223